• ATMOS LC 27 / ATMOS LC 27 Battery

    The homecare
    suction device
    for any situation

ATMOS LC 27 / ATMOS LC 27 Battery

The economical homecare suction device with a built-in guarantee of success

The ATMOS LC 27 homecare suction device was specifically designed to meet the requirements of bronchial suction in the homecare environment. It not only convinces affected patients, their relatives and caregivers, but also service providers in the medical field. Its technical sophistication makes the ATMOS LC 27 a worthwhile investment for all homecare providers and purchasing organisations in the care sector as well as for medical supply stores, pharmacies, and nursing homes. The ATMOS LC 27 meets all the user’s demands: it is self-explanatory, simple, immediately ready to use, safe, and portable.

ATMOS LC 27 Homecare Suction Device

Applications

  • Rest homes / nursing homes
  • Homecare sector

Intuitive operation

  • All controls required for operation are coloured blue. This feature aids users in performing suction.
  • The large ON/OFF button makes operation noticeably easier. The integrated LED lamp additionally indicates when the suction device is running.
  • The vacuum capacity is infinitely adjustable within a 180° range. It can be set to the patient’s individual needs, with the vacuum gauge showing easy-to-read values.

Intelligent connecting systems

  • The innovative ATMOS Direct Docking System (DDS) eliminates the need for a hose connection between the suction device and secretion canister. Instead, the ATMOS single-patient secretion canister, together with its lid, connects directly to the suction device.
  • The connecting system designed by ATMOS enables the user to connect the suction hose firmly to the secretion canister with a simple twist and to disconnect it just as easily after the suction process is completed.

Modern design

  • Thanks to its simple design and compact size, the ATMOS LC 27 can blend into almost any setting. It remains discreet, but is immediately at hand when you need it.

Quick reuse

  • DDS bacterial and viral filter cartridge: easy and quick to replace
  • Suction device: completely maintenance-free and easy to clean
  • Single-patient secretion canister: easy to change and usable over the long term

Detailed information for users

ATMOS has compiled key information on the new homecare suction device especially for patients, relatives, and caregivers.

ATMOS LC 27 Patient

Video



Versions

LC27_Zubehoer-01_160x160Px.jpg

ATMOS LC 27 / Serres®

LC27_Zubehoer-02_160x160Px.jpg

ATMOS LC 27 / Medi-Vac®

Technical Data

ATMOS LC 27ATMOS LC 27 Battery
Article number321.1000.0; 321.1300.0; 321.1200.0; 321.1100.0321.2000.0; 321.2300.0; 321.2100.0; 321.2200.0
Suction capacity27 l/min27 l/min
Max. vacuum–80 kPa
–800 mbar
–80 kPa
–800 mbar
Operating modeShort-term operation (30 min on. 30 min off)Short-term operation (30 min on. 30 min off)
Voltage100–240 V, 50/60 Hz100–240 V, 50/60 Hz
Dimensions (H x W x D)243 x 286 x 118 mm243 x 286 x 118 mm
Weight1.8 kg3.5 kg
Noise level< 65 dB(A)< 65 dB(A)
Battery type-Lead gel
Battery operation-1 h
Battery charging time-12 h

FAQ

How often does the DDS bacterial and viral filter cartridge have to be replaced?

The DDS bacterial and viral filter cartridge is a single-patient product. That means that the DDS bacterial and viral filter cartridge must be replaced when the device is changed to another patient. When the device is being used for the same patient, the DDS bacterial and viral filter cartridge has to be replaced every two months (provided the device is being used). Should the filter cartridge become contaminated or clogged with secretion, it must be replaced immediately.

How often does the suction hose have to be replaced?

The suction hose is a single-patient product. That means that the suction hose must be replaced when the device is changed to another patient. When the device is being used for the same patient, the suction hose has to be replaced every 4 weeks.

What does 'single-patient secretion canister' mean?

The secretion canister system of the ATMOS LC 27 is designed to be used for one patient only. When the device is being used for one patient, the secretion canister can be cleaned with commercially available cleaning agents — either by hand or simply in a dishwasher. When the device is changed to another patient, the homecare provider disposes of the secretion canister and replaces it with a new one.

How often can I boil the single-patient secretion canister?

The single-patient secretion canister system can be boiled a maximum of 30 times. After that, the single-patient secretion canister must be replaced.

Can I use the ATMOS LC 27 without the DDS bacterial and viral filter cartridge?

No. This special filter cartridge was designed to protect the device against oversuction and to protect patients / third parties from the danger of contamination. It is not possible to operate the device without the filter cartridge. For this reason, always keep at least one DDS bacterial and viral filter cartridge replacement ready.

How long can I use the ATMOS LC 27 Battery?

The battery operation time of the ATMOS LC 27 Battery is approximately 60 minutes (when the battery is fully charged).

Can the ATMOS LC 27 Battery also be used when the battery is discharged?

Yes. If the battery is completely discharged, the device can be operated on the mains.

Can the ATMOS LC 27 Battery be taken along in a plane with no problems?

The ATMOS LC 27 Battery can be taken along on plane trips in checked or carry-on baggage without difficulty. It is possible to operate the device during a flight. Individual instructions from the flight crew and airline-specific regulations must still be followed.

Downloads

Accessories

The handy carrying bag allows you to transport the ATMOS LC 27 and the consumables you need safely and conveniently.

Consisting of the DDS secretion canister (800 ml) and the DDS canister lid. For use with one patient only.

 

The hydrophobic bacterial and viral filter prevents the suction device from becoming contaminated and protects it against bacteria and viruses. Under normal conditions of use, the bacterial and viral filter cartridge offers effective protection for two months.

 

The ATMOS hose connecting system allows for easy connection of the hose. The clinically tested suction hoses from ATMOS can be used for up to four weeks if handled properly.

Contact

Are you interested in this product?